Modes of Ventilation
Conventional Ventilation
The modes of ventilation can be thought of as classifications based on how to control the ventilator breath. Traditionally ventilators were classified based on how they determined when to stop giving a breath. The three traditional categories of ventilators are listed below. As microprocessor technology is incorporated into ventilator design, the distinction among these types has become less clear as ventilators may use combinations of all of these modes as well as flow-sensing, which controls the ventilator breath based on the flow-rate of gas versus a specific volume, pressure, or time.
Breath Termination
- In a volume-cycled ventilator the ventilator delivers a preset volume of gas with each breath. Once the specified volume of breath is delivered, the positive pressure is terminated.
- In a pressure-cycled ventilator, once a preset pressure is reached within the ventilator, the breath is terminated.
- In a time-cycled ventilator, the termination of the breath occurs after a certain specified time period.
Both pressure and volume modes of ventilation have their respective limitations. Many manufacturers provide a mode or modes that utilize some functions of each. These modes are flow-variable, volume-targeted, pressure-regulated, time-limited modes (for example, pressure-regulated volume control - PRVC). This means that instead of providing an exact tidal volume each breath, a target volume is set and the ventilator will vary the inspiratory flow at each breath to achieve the target volume at the lowest possible peak pressure. The inspiratory time limits the length of the inspiratory cycle and therefore the I:E ratio. Pressure regulated modes such as PRVC or Auto-flow (Draeger) can most easily be thought of as turning a volume mode into a pressure mode with the added benefit of maintaining more control over tidal volume than with strictly pressure-control.
Breath Initiation
The other method of classifying mechanical ventilation is based on how to determine when to start giving a breath. Similar to the termination classification noted above, microprocessor control has resulted in a myriad of hybrid modes that combine features of the traditional classifications. Note that most of the timing initiation classifications below can be combined with any of the termination classifications listed above.
- Assist Control (AC). In this mode the ventilator provides a mechanical breath with either a preset tidal volume or peak pressure every time the patient initiates a breath. Traditional assist-control used only a preset tidal volume--when a preset peak pressure is used this is also sometimes termed Intermittent Positive Pressure Ventilation or IPPV. However the initiation timing is the same--both provide a ventilator breath with every patient effort. In most ventilators a back-up minimum breath rate can be set in the event that the patient becomes apneic. Although a maximum rate is not usually set, an alarm can be set if the ventilator cycles too frequently. This can alert that the patient is tachypneic or that the ventilator may be auto-cycling (a problem that results when the ventilator interprets fluctuations in the circuit due to the last breath termination as a new breath initiation attempt).
- Synchronized Intermittent Mandatory Ventilation (SIMV). In this mode the ventilator provides a preset mechanical breath (pressure or volume limited) every specified number of seconds (determined by dividing the respiratory rate into 60 - thus a respiratory rate of 12 results in a 5 second cycle time). Within that cycle time the ventilator waits for the patient to initiate a breath using either a pressure or flow sensor. When the ventilator senses the first patient breathing attempt within the cycle, it delivers the preset ventilator breath. If the patient fails to initiate a breath, the ventilator delivers a mechanical breath at the end of the breath cycle. Additional spontaneous breaths after the first one within the breath cycle do not trigger another SIMV breath. However, SIMV may be combined with pressure support (see below). SIMV is frequently employed as a method of decreasing ventilatory support (weaning) by turning down the rate, which requires the patient to take additional breaths beyond the SIMV triggered breath.
- Controlled Mechanical Ventilation (CMV). In this mode the ventilator provides a mechanical breath on a preset timing. Patient respiratory efforts are ignored. This is generally uncomfortable for children and adults who are conscious and is usually only used in an unconscious patient. It may also be used in infants who often quickly adapt their breathing pattern to the ventilator timing.
- Pressure Support Ventilation (PSV). This was developed as a method to decrease the work of breathing in-between ventilator mandated breaths. Thus, for example, SIMV might be combined with PSV so that additional breaths beyond the SIMV programmed breaths are supported. However, while the SIMV mandated breaths have a preset volume or peak pressure, the PSV breaths are designed to cut short when the inspiratory flow reaches a percentage of the peak inspiratory flow (e.g. 10-25%). Also, the peak pressure set for the PSV breaths is usually a lower pressure than that set for the SIMV breath. PSV can be also be used as an independent mode. However, since there is generally no back-up rate in PSV, appropriate apnea alarms must be set on the ventilator.
- Continuous Positive Airway Pressure (CPAP). A continuous level of elevated pressure is provided through the patient circuit to maintain adequate oxygenation, decrease the work of breathing, and decrease the work of the heart (such as in left-sided heart failure - CHF). Note that no cycling of ventilator pressures occurs and the patient must initiate all breaths. In addition, no additional pressure above the CPAP pressure is provided during those breaths. CPAP may be used invasively through an endotracheal tube or tracheostomy or non-invasively with a face mask or nasal prongs.
- Positive End Expiratory Pressure (PEEP) is functionally the same as CPAP, but refers to the use of an elevated pressure during the expiratory phase of the ventilatory cycle. After delivery of the set amount of breath by the ventilator, the patient then exhales passively. The volume of gas remaining in the lung after a normal expiration is termed the functional residual capacity (FRC). The FRC is primarily determined by the elastic qualities of the lung and the chest wall. In many lung diseases, the FRC is reduced due to collapse of the unstable alveoli, leading to a decreased surface area for gas exchange and intrapulmonary shunting (see above), with wasted oxygen inspired. Adding PEEP can reduce the work of breathing (at low levels) and help preserve FRC.
High Frequency Ventilation (HFV)
High-Frequency Ventilation refers to ventilation that occurs at rates significantly above that found in natural breathing (as high as 300-900 "breaths" per minute). Within the category of high-frequency ventilation, the two principal types are flow interruption and high-frequency oscillatory ventilation (HFOV). The former operates similarly to a conventional ventilator, providing increased circuit pressure during the inspiratory phase and dropping back to PEEP during the expiratory phase. In HFOV the pressure wave is driven by an electromagnetically controlled diaphragm similar to a loudspeaker. Because this can rapidly change the volume in the circuit, HFOV can produce a pressure that is lower than ambient pressure during the expiratory phase. This is sometimes called "active" expiration. In both types of high-frequency ventilation the pressure wave that is generated at the ventilator is markedly attenuated by passage down the endotracheal tube and the major conducting airways. This helps protect the alveoli from volutrauma that occurs with traditional positive pressure ventilation. Although the alveoli are kept at a relatively constant volume, similar to CPAP, other mechanisms of gas exchange allow ventilation (the removal of CO2) to occur without tidal volume exchange. Ventilation in HFV is a function of frequency, amplitude, and I:E ratio and is best described graphically as the area under the curve of an oscillatory cycle. Amplitude is analogous to tidal volume in conventional ventilation; larger amplitudes remove more CO2. Paradoxically, lower frequencies remove more CO2 in HFOV whereas in conventional ventilation the opposite is true. As frequency increases, the total time for a single cycle decreases (the oscillatory curve is shortened thereby decreasing the area under the curve and thus ventilation). I-time is set as a percentage of total time (usually 33%). Amplitude is a function of power and is subject to variability due to changes in compliance or resistance. Therefore, power requirements may vary significantly during treatment and from patient to patient. Patient characteristics and ventilator settings determine whether PaCO2 changes may be more sensitive to amplitude or frequency manipulation. In HFOV, mean airway pressure (MAP) is delivered via a continuous flow through the patient circuit which passes through a variable restriction valve (mushroom valve) on the expiratory limb. Increasing the flow through the circuit and/or increasing the pressure in the mushroom valve increases MAP. The MAP in HFOV functions similarly to PEEP in conventional ventilation in that it provides the pressure for alveolar recruitment.