Needle gauge comparison chart

Revision as of 22:05, 11 March 2009 by Zorkun (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

Hypodermic needles are available in a wide variety of outer diameters described by gauge numbers. Smaller gauge numbers indicate larger outer diameters. Inner diameter depends on both gauge and wall thickness. The following chart shows nominal inner diameter and wall thickness for regular-wall needles. Thin-wall needles (not shown) have identical outer diameters but larger inner diameters for a given gauge.

Needle Nominal Outer Diameter Nominal Inner Diameter
Gauge mm inches tol. (in.) mm inches tol. (in.)
10 3.404 0.1340 ±0.0010 2.692 0.1060 ±0.0020
11 3.048 0.1200 " 2.388 0.0940 "
12 2.769 0.1090 " 2.159 0.0850 "
13 2.413 0.0950 " 1.803 0.0710 "
14 2.108 0.0830 " 1.600 0.0630 "
15 1.829 0.0720 ±0.0005 1.372 0.0540 ±0.0015
16 1.651 0.0650 " 1.194 0.0470 "
17 1.473 0.0580 " 1.067 0.0420 "
18 1.270 0.0500 " 0.838 0.0330 "
19 1.067 0.0420 " 0.686 0.0270 "
20 0.902 0.0355 +0.0005

-0.0000

0.584 0.0230 +0.0015

-0.0000

21 0.813 0.0320 " 0.495 0.0195 "
22 0.711 0.0280 " 0.394 0.0155 "
22s 0.711 0.0280 " 0.140 0.0055 "
23 0.635 0.0250 " 0.318 0.0125 "
24 0.559 0.0220 " 0.292 0.0115 "
25 0.508 0.0200 " 0.241 0.0095 "
25s 0.508 0.0200 " 0.140 0.0055 "
26 0.457 0.0180 " 0.241 0.0095 "
26s 0.467 0.0184 " 0.114 0.0045 "
27 0.406 0.0160 " 0.191 0.0075 "
28 0.356 0.0140 " 0.165 0.0065 "
29 0.330 0.0130 " 0.165 0.0065 "
30 0.305 0.0120 " 0.140 0.0055 "
31 0.254 0.0100 " 0.114 0.0045 "
32 0.229 0.0090 " 0.089 0.0035 "
33 0.203 0.0080 " 0.089 0.0035 "

See also

External links

Template:WikiDoc Sources