Overview
Echocardiography is the best non-invasive test to evaluate the aortic valve anatomy and function. It is indicated in the case of presence of symptoms suggestive of valvular problems or in the case of detection of a systolic murmur with a grade greater than 3/6. Doppler echocardiography allows the measurement of the maximum jet velocity and can be used to estimate the effective orifice area of the aortic valve as well as the gradient across the aortic valve using the modified Bernoulli equation (gradient = 4 x velocity2). The flow must be constant, so as the velocity increases, the valve area decreases proportionally. Attention to technical details is important as they may lead to underestimation of the severity of the aortic stenosis[1]. Echocardiography can also be used to assess the severity of left ventricular hypertrophy.
Optimal Echocardiography Views
Parasternal Long Axis View
The parasternal long axis view visualizes the right and non coronary leaflets.
Parasternal Short Axis View
In the parasternal short axis view, the aortic leaflets open equally and form a circular orifice during systole. During diastole, the normal leaflets form a three pointed star with prominence at the closing point (nodules of Arentius).
Severity of Aortic Stenosis
Aortic stenosis severity can be assessed by estimating both the pressure gradient across the valve and the aortic valve area.
- Using echocardiographic flow velocities, the aortic valve area can be calculated non-invasively to estimate the severity of aortic stenosis.
- Using the velocity of the blood through the valve, the pressure gradient across the valve can be calculated by the equation:
Gradient = 4(velocity)² mmHg
Normal Aortic Valve:
- A normal aortic valve has no gradient, and
- Aortic valve surface area is of 2.5 to 3.5 cms2.
Mild Aortic Stenosis:
- Mean pressure gradient across the valve is less than 25 mm Hg, or
- Valve surface area is between 1.5 and 2.5 cms2, or
- Jet velocity less than 3.0 m per second
Moderate Aortic Stenosis:
- Mean pressure gradient across the valve is between 25 mm Hg and 40 mm Hg, or
- Valve surface area is between 1.0 and 1.5 cms2, or
- Jet velocity 3.0 to 4.0 m per second
Moderate to Severe Aortic Stenosis
- Valve surface area is between 0.7 and 1.0 cms2
Severe Aortic Stenosis:
- Mean pressure gradient across the valve is more than 40 mm Hg, or
- Valve surface area is less than 0.7 cm2, or
- Jet velocity is greater than 4.0 m per second
Severity | mild | moderate | severe |
---|---|---|---|
Valve Surface Area | 2.5 - 1.5 | 1.5 - 1.0 | <1.0 |
Jet Velocity (m/s) | 2 - 3 | 3 - 4 | >4 |
Peak Gradient (mmHg) | <25 | 25 - 40 | >40 |
Mean Gradient (mmHg) | <20 | 20-40 | >40 |
Low Flow, Low Gradient Aortic Stenosis
- Echocardiography is fairly accurate in the assessment of the severity of aortic stenosis in patients with normal or near normal cardiac output. However, dobutamine stress echocardiography or cardiac catheterization may be required to accurately assess the severity of aortic stenosis in patients with low output aortic stenosis.
- Low dose (<20 microgram/Kg/min) dobutamine stress echocardiography allows:
- The differentiation between severe and pseudo-severe low flow low gradient aortic stenosis through the degree of change of both the effective orifice area and the transvalvular gradient after dobutamine infusion. Changes following dobutamine infusion are due to the fact that dobutamine causes increase in the flow across the aortic valve.
- In pseudosevere aortic stenosis, there is increase in the effective orifice area of the aortic valve, in a way that it is larger than 1.2 cm2, and slight increase in the gradient, in a way that is inferior to 40 mmHg.
- In severe aortic stenosis, there is absence or little increase in the effective orifice area of the aortic valve, in a way that is inferior to 1.2 cm2, and increase in the gradient to become greater than 40 mmHg.
- The assessment of the presence or absence of left ventricular flow reserve through the degree of change in stroke volume following dobutamine infusion.
- Left ventricular flow reserve is an indicator of operative mortality. Left ventricle flow reserve does not predict recovery of left ventricle after aortic valve replacement nor the resolution of symptoms.[2]
- The aortic valve area should increase to more than 1-1.2 cm2 with a dobutamine infusion and the mean pressure gradient should rise above 30 mm Hg. While early surgical mortality is 32-33% in patients who fail to to achieve these improvements with dobutamine infusions, it is only 5–7% in patients who demonstrate an increase in their contractility and gradient. Five years survival after surgery was 88% in patients with improved contractility and only 10–25% in patients with unimproved contractility.
- The differentiation between severe and pseudo-severe low flow low gradient aortic stenosis through the degree of change of both the effective orifice area and the transvalvular gradient after dobutamine infusion. Changes following dobutamine infusion are due to the fact that dobutamine causes increase in the flow across the aortic valve.
- Shown below is an algorithm summarizing the interpretation of the results of low dose dobutamine stress echocardiography in the evaluation of low flow, low gradient aortic stenosis.
Examples
Demonstration of Diastolic Mitral Regurgitation due to Severe Aortic Stenosis
Demonstration of Aortic regurgitation combined with Aortic Stenosis
{{#ev:youtube|J-Qs_evodww}}
Demonstration of Calcific Aortic Stenosis with Mitral Annular Calcification and Moderate Mitral Regurgitation
{{#ev:youtube|qjpV19034sk}}
Low flow Aortic Stenosis
{{#ev:youtube|WC3uKS4ShSM&feature=related}}
Severe Aortic Stenosis without Aortic Regurgitation
Cntinuous wave doppler below shows a peak velocity across the aortic valve of a gradient of 85 mmHg, which is compatible with severe aortic stenosis. No aortic regurgitation is present.
2008 and Incorporated 2006 ACC/AHA Guidelines for the Management of Patients with Valvular Heart Disease (DO NOT EDIT) [1]
Echocardiography (DO NOT EDIT) [1]
Class I |
"1. Echocardiography is recommended for the diagnosis and assessment of aortic stenosis severity.(Level of Evidence: B) " |
"2. Echocardiography is recommended in patients with aortic stenosis for the assessment of left ventricular wall thickness, size, and function.(Level of Evidence: B) " |
"3. Echocardiography is recommended for re-evaluation of patients with known aortic stenosis and changing symptoms or signs.(Level of Evidence: B) " |
"4. Echocardiography is recommended for the assessment of changes in hemodynamic severity and left ventricular function in patients with known aortic stenosis during pregnancy. (Level of Evidence: B) " |
"5. Transthoracic echocardiography is recommended for re-evaluation of asymptomatic patients: every year for severe aortic stenosis; every 1 to 2 years for moderate aortic stenosis; and every 3 to 5 years for mild aortic stenosis.(Level of Evidence: B) " |
Class IIa |
"1. Dobutamine stress echocardiography is reasonable to evaluate patients with low-flow/low-gradient AS and LV dysfunction. (Level of Evidence: B)" |
Asymptomatic Adolescents (DO NOT EDIT) [1]
Class I |
"1. Doppler echocardiography is recommended yearly in the asymptomatic adolescent or young adult with AS who has a Doppler mean gradient greater than 30 mm Hg or a peak velocity greater than 3.5 m per second (peak gradient greater than 50 mm Hg) and every 2 years if the Doppler gradient is less than or equal to 30 mm Hg or the peak jet velocity is less than or equal to 3.5 m per second (peak gradient less than or equal to 50 mm Hg).(Level of Evidence: C) " |
Sources
- 2008 Focused Update Incorporated Into the ACC/AHA 2006 Guidelines for the Management of Patients With Valvular Heart Disease [1]
References
- ↑ 1.0 1.1 1.2 1.3 1.4 Bonow RO, Carabello BA, Chatterjee K, de Leon AC, Faxon DP, Freed MD; et al. (2008). "2008 Focused update incorporated into the ACC/AHA 2006 guidelines for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the 1998 Guidelines for the Management of Patients With Valvular Heart Disease): endorsed by the Society of Cardiovascular Anesthesiologists, Society for Cardiovascular Angiography and Interventions, and Society of Thoracic Surgeons". Circulation. 118 (15): e523–661. doi:10.1161/CIRCULATIONAHA.108.190748. PMID 18820172.